Arkray PocketChem BA PA-4140
POCT Blood Ammonia (N-NH₃) Monitoring System for Inborn Errors of Metabolism and Hyperammonemia
REF: 107687| SKU: 500387| BRAND: Arkray
🎯
Reflectance Dry Chemistry (LED 635 nm)
🧪
20 µL Capillary or Venous Whole Blood
⏱️
Approx. 3 min 20 sec
🏥
Emergency Room, Intensive Care, Pediatrics, POCT
Scientific Description and IVD Diagnostic Value
The Arkray PocketChem BA PA-4140 blood ammonia analyzer is an essential In Vitro Diagnostic (IVD) Point of Care (POCT) device for the rapid quantitative determination of ammonium (N-NH₃) in whole blood. Timely determination of blood ammonia is a critical parameter in the management of clinical emergencies, hospital emergency department diagnostics, and hospital ward rapid diagnostics (Pediatrics, Intensive Care, Neonatology, and Rare Metabolic Disorders).
The system is essential for monitoring and follow-up of patients suffering from complex pathologies such as Urea Cycle Disorders (UCD) and other forms of inherited hyperammonemia / inborn errors of metabolism. Specifically, the PocketChem BA is a valuable aid in evaluating and managing:
- Ornithine Transcarbamylase (OTC) Deficiency: the most common genetic form among urea cycle defects, characterized by the risk of acute decompensating hyperammonemic crises.
- Carbamoyl Phosphate Synthetase (CPS1) Deficiency: a severe metabolic condition with often neonatal onset.
- Citrullinemia (Type I and II): an inborn error responsible for toxic ammonium accumulation and metabolic encephalopathy.
- HHH Syndrome (Hyperornithinemia-Hyperammonemia-Homocitrullinuria): autosomal recessive pathology of ornithine transport with urea cycle alteration.
- Secondary Hyperammonemia: acute liver failure, hepatic encephalopathy, and metabolic decompensation in emergency settings.
Operating principle and advantages over traditional systems:
Traditional laboratory ammonia analysis requires stationary venous sampling, immediate ice placement, rapid refrigerated centrifugation, and execution within very few minutes to prevent in vitro amino acid deamination, the primary source of false positives. The Arkray PocketChem BA PA-4140 overcomes these limitations by operating directly on a micro-sample of only 20 µL of whole blood (from fingertip or earlobe capillary puncture), without any need for centrifugation or pre-treatment. Utilizing the membrane microdiffusion method combined with dry chemistry (Ammonia Test Kit II), the system separates the gaseous phase of ammonia and measures colorimetric variation via 635 nm reflection photometry, ensuring high diagnostic accuracy while eliminating waiting times and pre-analytical flaws.
Operational Advantages for Laboratory and Ward Instrumentation
- Rapid decision-making: Clinical results available in just 3 minutes and 20 seconds, ideal for rapid medical ward and Intensive Care diagnostics.
- Drastic reduction of pre-analytical errors: By eliminating centrifugation and ice storage of the sample, risks of hemolysis and artificial in vitro ammonia increases are nullified.
- Micro-sample volume (20 µL): Volumetric requirement kept to a minimum, perfect for neonatology and pediatric applications.
- Simplicity and Automation: Intuitive procedure with calibration embedded directly in the Ammonia Test Kit II test strips via barcode/chip.
- Integration into the clinical workflow: Extremely compact dimensions (150 g) with a detachable thermal printer and RS-232C serial interface for PC/LIS connection.
Step-by-Step Operational Workflow
- Sample Preparation: Perform the capillary puncture (finger or earlobe) or complete blood draw and aspirate exactly 20 µL of whole blood using a pipette or capillary tube.
- Insertion into the Instrument: Deposit it onto the well of the Ammonia Test Kit II test strip housed in the tray and press the START button.
- Automatic Processing: The instrument starts the reaction countdown (180 seconds) maintaining integrated thermal control. An acoustic signal alerts upon reaction completion.
- Result Interpretation: Remove the spacer tab of the strip, reposition it on the tray, and immediately view the ammonium value expressed in N-µg/dL (or µmol/L) on the LCD display, with automatic report printing.
Device Technical Specifications
| Technical Parameter | IVD Device Specifications |
|---|---|
| Analyzed Parameter | Blood Ammonia / N-NH₃ |
| Analytical Methodology | Single-wavelength reflectance dry chemistry (LED 635 nm) with membrane microdiffusion |
| Measuring Range | 10 – 400 N-µg/dL (8 – 285 µmol/L) |
| Sample Type & Volume | 20 µL of whole blood (Capillary or Venous) |
| Analytical Time | Approx. 3 minutes and 20 seconds per test (Pre-heating 1-2 min) |
| Test Strip & Calibration | Ammonia Test Kit II (Calibration curve pre-stored in kit) |
| Thermal Compensation | Automatic via integrated temperature sensor (Range 10 – 35 °C) |
| Memory & LIS Connectivity | 50 measurements; RS-232C serial interface for data export to PC/LIS |
| Dimensions & Weight | 124 (W) x 85 (D) x 38 (H) mm | Weight: approx. 150 g (without batteries) |
| Power Supply | 2 AA alkaline batteries (1.5V) and/or AC power adapter (required for printer use) |
| Thermal Printer (Optional module) | Integrated/detachable contact unit (180 g, paper width 58 mm) |
Technical Opinion: “The Arkray PocketChem BA PA-4140 analyzer represents a gold standard in POCT blood ammonia diagnostics. An indispensable IVD solution both for managing urgent care at the Hospital Emergency Department and Intensive Care, and for Rare Metabolic Disease centers dedicated to monitoring patients with inborn errors of the urea cycle (UCD, OTC deficiency, CPS1, citrullinemia, and HHH). The ability to deliver an accurate analytical result in just over 3 minutes from a whole blood micro-sample eliminates critical issues related to pre-analytical sample decay. Biolab Srl guarantees reagent supply, analytical validation support, and dedicated technical assistance.”