Fluorecare CK-MB (Creatine Kinase Isoenzyme)
REF: MF-32 | SKU: 113803 | BRAND: MicroProfit
⚙️
FIA – Fluorescence
🧪
Serum / Plasma
⏱️
15 Minutes
📦
20 Test Kit
Scientific Description
The fluorecare CK-MB diagnostic kit is an advanced diagnostic aid designed for the rapid and quantitative determination of Creatine Kinase Isoenzyme (CK-MB) concentration in human serum or plasma samples. The system is dedicated exclusively for use and quantitative analysis in combination with the Microprofit Fluorecare Reader dry immunofluorescence analyzer (model MF-T1000).
The analytical principle is based on a double-antibody sandwich immunochromatographic assay: the human serum or plasma specimen is loaded into the sample well; CK-MB present in the biological sample combines with the fluorescence-labeled anti-CK-MB antibody located on the bonding pad, forming the corresponding antibody complex. Driven by capillary action, the complex migrates along the nitrocellulose membrane to the test line (T), where it binds to the pre-coated capture antibody in the detection zone. A higher concentration of CK-MB in the sample results in a more substantial accumulation of the complex on the test line, increasing the intensity of the fluorescent band under a specific wavelength. The emitted signal intensity is directly proportional to the analyte concentration in the biological sample. The analyzer quantifies optical density and converts the signal into an analytical value expressed in ng/mL, based on a pre-set calibration curve provided via a dedicated ID Chip. The test is intended solely for professional in vitro use as a clinical detection support in Human Clinical Analysis Laboratories, Clinics, and Nursing Homes.
Execution Protocol
- Equilibrate the test card and specimens to room temperature (20-25°C) prior to starting procedures.
- Power on the analyzer following the instrument operating instructions carefully.
- Verify exact matching between the lot number on the ID Chip and that on the test cards included in the diagnostic kit, then insert the ID Chip avoiding touching the gold connector contacts.
- Open the aluminum foil pouch by tearing along the pre-cut notch, take out the test card and place it on a flat work surface; perform the analytical assay within 1 hour of opening.
- Holding the pipette vertically without generating air bubbles, dispense exactly 70 µL of sample (serum or plasma) or reference solution directly into the sample well of the test card.
- Allow the card to incubate at room temperature for a static period of 15 full minutes.
- At the end of the incubation time, insert the card into the holder slot of the Fluorecare analyzer and launch the reading cycle; quantitative results (expressed in ng/mL) will display on screen within seconds, ready for printing or exporting.
Performance Table
| Performance Parameter | Value / Technical Specification |
|---|---|
| Linearity Range (Dynamic Range) | 1.0 – 80 ng/mL (R > 0.975) |
| Analytical Sensitivity (LOD) | ≤ 1.0 ng/mL |
| Accuracy (Relative Deviation) | Within ±15% |
| Intra-assay Precision (Within-run CV) | ≤ 15% |
| Inter-assay Precision (Between-run CV) | ≤ 15% |
| Suggested Reference Range | 0 – 5 ng/mL |
Technical Opinion: The MicroProfit Fluorecare quantitative immunofluorescence system ensures rigorous linear correlation in monitoring the myocardial biomarker CK-MB. The streamlined analytical workflow and native calibration standardized via ID Chip minimize operational margins of error in the field, providing robust emergency clinical detection support directly within the medical facility. For professional in vitro diagnostic use only. Supply strictly reserved for Clinical Analysis Laboratories, Clinics, and Nursing Homes.
Technical Team BIOLAB Diagnostici – Laboratory Division
Biolab is the exclusive Italian distributor of MicroProfit